Steam Sterilizer Horizontal Cylindrical

Sterilizer Equipments   |    Steam Sterilizer Horizontal Cylindrical
Steam Sterilizer Horizontal Cylindrical

Our horizontal Steam Sterilizer are widely used in various bulk laboratories/Hospital for their day to day studies on scale sterilization procedures where they give dryness of the sterilized medium a paramount importance and for that our sterilized media is prone to various micro-bacterial and viral infections. Having a specialization in customer's requirements we have an expert team of making both standard and customize models.





Sailent Features
  • Cylindrical design triple walled horizontal autoclave mounted on a sturdy, heavy M.S. duly painted tubular stand.
  • Inner chamber, door lid, jacket, boiler & outer jacket, boiler & outer jacket cover made of SS304.
  • Steam is generated in a separate boiler which is placed below the mail body.
  • Lid is provided with pressure locking system which prevents opening of the door while Steam is inside the chamber.
  • Single hinged door and manual radial locking system made of mild steel nickel coated industrial grade energy efficient heaters reduce power bills drastically
  • Gap between Jacket and outer cover wall is filled with high grade fiber/mineral glass wool insulation.
  • Fitted with pressure gauge for jacket, compound gauge for chamber, spring loaded safety valve and steam release valve.
  • Automatic Vacuum breaker is provided to break the vacuum in case of formation of vacuum due to steam condensation.
  • Automatic Pressure Control Switch for additional safety. When steam pressure inside jacket reaches 18 to 20 psi it automatically cuts off electric supply to heater.
  • Automatic Low Water Level cut off device fitted for
  • general safety of the element.
  • Digital PID temperature indicator cum controller- Controller has LED display which displays set value and process value. It has inbuilt automatic digital minute timer which start automatically when set/desired value of temperature is achieved. When the set time period is over heaters get switched off automatically with alarm for cycle end.
  • Manually operated multiport valve having 4 positions is provided for steam injection inside the chamber, dry sterilization & steam exhaust.
  • For ensuring proper sealing of the door with the chamber and zero leakage a high tensile joint less Silicon
  • gasket is used which is seated in an integrated groove.
  • SS Baffle plate for effective steam distribution and avoid direct hitting of steam on sample load
  • Validation port provided
  • Chamber Condensate Line is incorporated with steam trap for perfect condensation to get optimum temperature.
  • Fitted with plug screen to prevent line choking due to sediment discharge
  • SS bottom plate in chamber to avoid the load from coming in contact with condensate
  • Autoclave is fitted with water level indicator glass gauge and easily replaceable heaters
  • Chamber and Jacket is made of 14 SWG.
  • Hydraulically tested at 40psi 2.5 times the operating pressure of autoclave
  • Small sized autoclave works on 230V single phase power supply and big size works effiently on 440 V three phase power supply